Master's thesis of Applied Science: Image quality and radiation dose comparison of a computed radiography system and an amorphous silicon flat panel system in paediatric radiography: a phantom study
lượt xem 3
download
This purpose of this work was to investigate the patient radiation doses and image quality of a Philips/Agfa computed radiographic (CR) system and a Philips indirect-capture digital radiographic (DR) system in a paediatric setting. A CDRAD digital radiographic contrast-detail phantom was used to assess radiographic image quality. Perspex slabs of three different thicknesses (6, 11 and 16 cm) were used to simulate paediatric patients of three arbitrary ages. These phantoms, in conjunction with the CDRAD digital radiographic contrast-detail phantom, were imaged under three different conditions. The CDRAD Analyser software package was used to assess the quality of each image.
Bình luận(0) Đăng nhập để gửi bình luận!
CÓ THỂ BẠN MUỐN DOWNLOAD